Features: Baud rates to 230,400
High speed to 500 readings/second
Lower cost "Industrial Stability" option
±15kV ESD protected
Real mV/V calibration
Extreme noise immunity 5 x heavy industrial level
Transducer calibration
Remote shunt calibration
Very high stability
Peak and trough recording
Programmable dynamic filter
Wide operating voltage (5.4V - 18V)
DC Excitation for longer cable lengths
Diagnostics LED
Advanced Characteristics:
An Intelligent Strain Transducer
High Performance
3 Year Manufacturer's Warranty
Serial Communications - excellent stability, RS232 and RS485 versions
Fieldbus connectivity - universal systems compatibility
Direct output in Engineering Units if required
Resolution selectable to 1 part in 400,000
"Plug in and go" sensor - no need for a separate amplifier
No need for an expensive A-to-D Card
Outstanding performance-to-cost ratio
Five-point transducer temperature compensation via comms
Seven-point transducer linearisation via comms
Sensor operation limit alarms
Reverse polarity protected
Continuous auto-zero operation
On-board temperature measurement as option
Optionally, temperature may be measured with an external probe
RS232 versions may be programmed to "stream" data continuously
Ideal for high precision measurement at low cost
Non-expert maintenance
Capable of ATEX Approval
Self diagnostics alerts user to common Strain Gauge faults:
Excitation short-circuits
Input over- or under-range
Cell scaling over- or under-range
System scaling over- or under-range
Temperature over- or under-range
Communication errors
Small package size
Works with most common strain gauge devices
Saves cost and space
All you need to add is a DC power supply
Simple to use
Competitive prices with generous quantity discounts
Ideal for OEM, R&D, Production, QA, QC applications
Can connect up to 254 devices on one RS485 bus
Full CE approval
Options:
Temperature compensation via external sensor
OEM Application Software through Bootloader,
allowing special functions to be programmed via RS485
RS232 - RS485 Converter
Alternative sensitivities (1 - 20 mV/V)
Precision vs Speed of Reading:
Generally, the higher the precision required from an analogue-to-digital system, the slower the speed of conversion to digital format. The DSC allows you to select the resolution and speed of reading that you require according to the Table below:
Speed
Readings/sec
(Hz) |
Resolution |
|
DSCH Models
High Stability |
DSCS Models
Industrial Stability |
|
|
ppm |
% of
Full Scale |
Divisions
1 part in... |
ppm |
% of
Full Scale |
|
1 |
400,000 |
2.5 |
0.00025 |
100,000 |
10 |
0.0010 |
|
10 |
120,000 |
8.3 |
0.00083 |
40,000 |
25 |
0.0025 |
|
100 |
50,000 |
20 |
0.0020 |
10,000 |
100 |
0.0010 |
|
500 |
18,000 |
56 |
0.0056 |
5,000 |
200 |
0.0500 |
Choice of Communication Options:
| RS232: |
ASCII Protocol - Data Streaming possible |
| . |
MANTRABUS Protocol |
| . |
MODBUS Protocol |
| RS485: |
ASCII Protocol |
| . |
MANTRABUS Protocol |
| . |
MODBUS Protocol |
| CAN Versions: |
MantraCAN Protocol
CANOpen Protocol |
Product Order Codes:
|
. |
High Stability DSCH |
Industrial Stability DSCS |
|
RS232 ASCII Protocol |
DSCH2ASC |
DSCS2ASC |
|
RS232 MANTRABUS II Protocol |
DSCH2MAN |
DSCS2MAN |
|
RS232 Modbus RTU Protocol |
DSCH2MOD |
DSCS2MOD |
|
RS485 ASCII Protocol |
DSCH4ASC |
DSCS4ASC |
|
RS485 MANTRABUS II Protocol |
DSCH4MAN |
DSCS4MAN |
|
RS485 Modbus RTU Protocol |
DSCH4MOD |
DSCS4MOD |
|
MantraCAN Protocol |
DSCHMCAN |
DSCSMCAN |
|
CANOpen Protocol |
DSCHCOP |
DSCSCOP |
Accessories:
DSJ1 Junction box for easy mounting of one DSC, either RS232 or RS485
DSJ4 Junction box for easy mounting of up to 4 DSCs, RS485 versions only
DS485DIS RS485 digital LED data display module
DSC Evaluation Kits
RS232 - RS485 Converter
VisualLink Rapid Development SCADA software
Specifications:
| Description |
Units |
High Stability |
Industrial Stability |
| Min |
Typ |
Max |
Min |
Typ |
Max |
| Bridge Excitation: |
VDC |
4.5 |
5 |
5.25 |
4.5 |
5 |
5.25 |
| Bridge Impedence: |
Ohms |
320 |
350 |
5,000 |
320 |
350 |
5,000 |
| Bridge Sensitivity: |
mV/V |
-3 |
- |
+3 |
-3 |
- |
+3 |
| Offset Temperature Stability: |
ppm/°C |
- |
1 |
4 |
- |
5 |
10 |
| Gain Temperature Stability: |
ppm/°C |
- |
3 |
5 |
- |
30 |
50 |
| Offset Stability With Time: |
% FR |
- |
0.002 |
0.008 |
- |
0.0035 |
0.016 |
| Gain Stability With Time: |
ppm of FR/1st Year |
- |
- |
30 |
- |
- |
300 |
| Nonlinearity: |
% FR |
- |
0.0005 |
0.0025 |
- |
0.0005 |
0.0025 |
| Internal Resolution: |
Counts/Divisions |
- |
16 Million |
- |
- |
16 Million |
- |
|
Resolution
@ 1 Hz: |
Counts/Divisions over 100s |
- |
400,000 |
- |
- |
100,000 |
- |
Resolution
@ 10 Hz: |
Counts/Divisions over 100s |
- |
120,000 |
- |
- |
40,000 |
- |
Resolution
@ 100 Hz: |
Counts/Divisions over 100s |
- |
50,000 |
- |
- |
10,000 |
- |
Resolution
@ 500 Hz: |
Counts/Divisions over 100s |
- |
18,000 |
- |
- |
5,000 |
- |
|
Optional:
Temperature Measurement Resolution |
°C |
- |
0.1 |
- |
- |
Not available |
- |
|
Optional:
Temperature Measurement Accuracy |
°C |
- |
1 |
- |
- |
Not available |
- |
| Power Supply Voltage: |
VDC |
5.4 |
12 |
18 |
5.4 |
12 |
18 |
| Power Supply Noise/Ripple: |
|
- |
- |
100 |
- |
- |
100 |
| Power Supply Current (350R Bridge): |
mA |
- |
45 |
60 |
- |
45 |
60 |
| Power @ 10V Supply (350R Bridge): |
mW |
- |
350 |
- |
- |
350 |
- |
| Sensor Impedence: |
Ohms |
320 |
350 |
5,000 |
320 |
350 |
5,000 |
|
Excitation System: |
- |
- |
4-wire |
- |
- |
4-wire |
- |
| Operating Temperature Range: |
°C |
-40 |
- |
+85 |
-40 |
- |
+85 |
| Operating Temperature Range for OIML 6000d: |
°C |
-10 |
- |
+55 |
- |
- |
- |
| Storage Temperature: |
°C |
-40 |
- |
+85 |
-40 |
- |
+85 |
| Humidity: |
% RH Non-condensing |
0 |
- |
95 |
0 |
- |
95 |
| RS485 Data Rate: |
Baud |
2,400 |
- |
230,400 |
2,400 |
- |
230,400 |
| CAN Bit Rate: |
Bits/s |
10k |
- |
1M |
10k |
- |
1M |
| Package: |
PCB with metal shield, IP40 protected |
| Dimensions: |
Length 87 x Width 20 x Height 11.75 (mm) |
| Environmental Approvals: |
EMC Directive 89/336/EEC
EMC Emissions
EN 50 081-1 :1992 (Light Industrial)
EN 50 081-2 :1992 (Heavy Industrial)
EMC Immunity
EN 50 082-1 :1992 (Light Industrial)
(RF Field Test 0.001% FS over 30 to 500MHz)
EN 50 082-2 :1992 (Heavy Industrial)
(RF Field Test 0.005% FS over 30 to 500MHz)
Low Voltage Directive 73/23/EEC amended by 93/68/EEC
IEC 1010-1 :1990
BSEN 610101 :1993 |
Buy securely OnLine now using

Please select the appropriate shipping method
then click "Add To Cart"
HIGH STABILITY OPTIONS
WITH RS485 OR CAN COMMUNICATIONS
HIGH STABILITY OPTIONS
WITH RS232 COMMUNICATIONS
INDUSTRIAL STABILITY OPTIONS
WITH RS485 OR CAN COMMUNICATIONS
INDUSTRIAL STABILITY OPTIONS
WITH RS232 COMMUNICATIONS
EASY-CONNECT MODULES
& ACCESSORIES
Please contact us for quantity prices etc
|

DSC Digital Strain Card

Free Instrument Explorer software allow you to configure and run your DSC

DCSLOG24 Freeware Software allows up to 24 DSCs and/or Dcell channels to be displayed and logged simultaneously

DSJ1 Easy-connect module for 1 DSC (RS232 or RS485 versions)

DSJ4 Easy-connect module for up to 4 DSC cards (RS485 & CAN versions only)

DS485DIS Digital Display Module for up to 8 DSCs or DCells

DSCUSB USB version of the DSC, with DSCUSB Toolkit software
Buy securely OnLine now using

Other products that may be of interest:
T24 Wireless Radio
Telemetry System
SGA Analogue Strain
Gauge Amplifier
ICA In-Cell Amplifier Modules
ADP15 Process Controllers
ADW15 Weighing Controllers
LVDT & RVDT Conditioning
Amplifier
LCA15 Intelligent Controllers
SMP Surface-Mount Controllers
RCA15 Rack-Mount Amplifier
UAB Rack Mount Amplifier
Load Cells
Accelerometers
Please email us if you need any more information
|